Load Default MPC Parameter Values

The Afimilk MPC is supplied with standard parameter values that are pre-loaded in its operating system. In some countries, local parameter values are used. The standard value of the parameter Load is 0. If you are using different set of parameters, use the local load number provided by Afimilk.

There are two procedures for loading preset parameters values:

  • Full load – replaces all the parameter values stored in the control unit, including stall ID number and all calibration parameters.

    Note: Carry out "Full Load" procedure at system startup OR when replacing a faulty display.

  • Partial load – replaces operational parameters with new values, preserving existing calibration values and stall ID numbers.

    Note: Use this procedure if a parameter load at operating parlor level is needed

The MPC parameters are set and updated via AfiFarm. Default parameters are initially loaded to the MPC. However, default parameters can differ between countries. Afimilk provides several options for loading.

Note: Ask your local dealer for your country "load" number OR use Load No. 0 as explained below.

To load default parameters

  1. Make sure the milk meter is in wash mode:

  2. Use the Cancel ACR to select a partial or full load procedure, as follows:

    • LED ON = Full load is indicated by an illumated LED indicator

    • LED OFF = Partial load is indicated when the LED is off

  3. Press .

    appears on the display.

    (00 represents the number of the set

  4. Type your regional Load number (if one is delivered by your dealer).

  5. Press to start the loading process.

    Note: Press to discard changes and exit programming mode.